Why does CANNON offer so many different types of glass capillary viscometers? Primarily because no single capillary viscometer is ideally suited for all kinematic viscosity determinations. Individual analysts also have their own preferences. The tables on the right may assist you in determining which viscometer is most appropriate for your particular application.

Expanded Uncertainty

CANNON Instrument Company provides expanded uncertainty data with all of our calibrated viscometers. This is the universally accepted statistic when dealing with calibration data. In order to maintain A2LA accreditation, we are required to determine and specify the expanded uncertainty for all our calibration data. ISO publications define expanded uncertainty as "quantity defining the interval about the result of a measurement within which the values that could reasonably be attributed to the measurement may be expected to lie with a high level of confidence."

Expanded uncertainty is a more comprehensive and meaningful term than "precision" because it incorporates all variables in the measurement process. When comparing the accuracy of viscometers it is essential that the expanded uncertainty be specified. Detailed information regarding CANNON measurements and expanded uncertainty is available for download from the CANNON website, www.cannoninstrument.com.

CANNON Instrument Company also provides uncertainty data with every bottle of our certified viscosity standards.

First Name
required
Last Name
required
Email
required
Telephone
 
Enquiry
 
Allow contact by email
 

Call for current prices

Cannon® Glass Capillary Viscometers
Technical data for:
Cannon-Fenske, Cannon-Ubbelohde and Cannon-Ubbelohde Dilution Viscometers
 
 
Size
Approx
constant,
cSt/s
 
Range
centistokes*
25
0.002
0.5 to 2
50
0.004
0.8 to 4
75
0.008
1.6 to 8
100
0.015
3 to 15
150
0.035
7 to 35
200
0.1
20 to 100
0.25
0.25
50 to 250
350
0.5
100 to 500
400
1.2
240 to 1200
450
2.5
500 to 2500
500
8
1600 to 8000
600
20
4000 to 20000
650
45
9000 to 45000
700
100
20000 to 100000

* Ranges in centistokes represent flow times of 250 to 1000 seconds for Size 25, and 200 to 1000 for all other sizes.

Technical data for:
Ubbelohde Viscometers
 
 
Size
Approx
constant,
cSt/s
 
Range
centistokes*
0
0.001
0.3 to 1
0C
0.003
0.6 to 3
0B
0.005
1 to 5
1
0.01
2 to 10
1C
0.03
6 to 30
1B
0.5
10 to 50
2
0.1
20 to 100
2C
0.3
60 to 300
2B
0.5
100 to 500
3
1.0
200 to 1000
3C
3.0
600 to 3000
3B
5.0
1000 to 5000
4
10
2000 to 10000
4C
30
6000 to 30000
4B
50
10000 to 50000
5
100
20000 to 100000