C 2000

The C 2000 basic IKAź calorimeter is a combustion calorimeter for determining gross calorific values of liquid and solid samples. A high level of automation with extremely simple handling characterizes this instrument. In addition to the isoperibolic measurement procedure (static jacket), a dynamic (reduced-time) working method is also available. To provide the calorimeter with cooling water, it needs to be connected to a thermostat f.e. IKA KV 600 ( accessory) or a firmly installed water connection. The C 2000 basic is equipped with a very convenient console to operate the unit.

  • Automatic water handling system includes tempering, filling and emptying of calorimeter inner vessel
  • Automatic oxygen filling of decomposition vessel
  • Automatic decomposition vessel identification
  • Automatic sample ignition
  • Validation according to DIN 51900,ISO 1928, ASTM D240, ASTM D4809, ASTM D5865, ASTM D1989, ASTM D5468, ASTM E711
  • Working methods:
    • isoperibol, measurement time: approx. 22 min
    • dynamic, measurement time: approx. 7 min
  • Compact, integrated modular design for convenient operation
  • Cooling water supply via thermostat f.e. IKA KV 600 (accessory) or firmly installed water connection (C 25 pressure regulating valve recommended)
  • Interface connections for each of the following: scale, printer, monitor and sample rack C 5020
  • User-friendly software C 5040 CalWin for controling the calorimeter and administration of measuring data
  • Up to eight measurement cells can be controlled by a single PC, using a multi-serial plug-in card PCI 8.2 (accessory)
  • LIMS integration is possible
  • Special halogen resistent vessel for quantitative decomposition of halogens and sulphur (accessory)
  • The decomposition vessel can be changed over to use combustible crucibles C 14 (accessory C5010.4 is needed)
  • Consumables for calibrations and initial operation are included with delivery.

C 2000 basic Version 1

Consisting of:

  • C 2000 basic
  • C 5010 Decomposition vessel, standard

C 2000 basic Version 2

Consisting of:

  • C 2000 basic
  • C 5012 Decomposition vessel, halogen resistant

C 2000 basic high pressure

Consisting of:

  • C 2000 basic
  • C 62 Decomposition vessel, high pressure
  • C 60 Conversion set

C 2000 control Version 1

Consisting of:

  • C 2000 control
  • C 5010 Decomposition vessel, standard
  • C 5040 CalWin, calorimeter software

C 2000 control Version 2

Consisting of:

  • C 2000 control
  • C 5012 Decomposition vessel, halogen resistant
  • C 5040 CalWin, calorimeter software

C 2000 control high pressure

Consisting of:

  • C 2000 basic
  • C 62 Decomposition vessel, high pressure
  • C 60 Conversion set
First Name
required
Last Name
required
Email
required
Telephone
 
Enquiry
 
Allow contact by email
 

Call for current price

C 2000
Technical data
Measuring range max.
40000 J
Measuring mode adiabatic 22°C
no
Measuring mode isoperibol 22°C
no
Measuring mode dynamic 25°C
yes
Measuring mode isoperibol 25°C
yes
Measuring mode dynamic 30°C
yes
Measuring mode isoperibol 30°C
yes
Measuring mode double dry (ISO 1928)
no
Measuring time dynamic approx
7 min
Measuring time isoperibol approx
22 min
Reproducibility dynamic
(1g benzoic acid NBS39i)
0.1 %RSD
Reproducibility isoperibol
(1g benzoic acid NBS39i)
0.05 %RSD
Working temperature max.
30 °C
Temperature measurement resolution
0.0001 K
Cooling medium temperature min.
12 °C
Cooling medium temperature max.
28 °C
Cooling medium permissible operating pressure
1.5 bar
Cooling medium
tap water
Type of cooling
flow
Interface scale
RS232
Interface printer
Centronix
Interface PC
RS232
Interface test rack
yes
Interface ext. monitor
yes
Interface ext. keyboard
yes
Oxygen filling
yes
Degasification
no
Decomposition detection
yes
Decomposition vessel C 5010
yes
Decomposition vessel C 5012
no
Decomposition vessel C 7010
no
Decomposition vessel C 7012
no
Decomposition vessel C 62
no
Analysis according to DIN 51900 (1977/84)
yes